Elettra-Sincrotrone Trieste S.C.p.A. website uses session cookies which are required for users to navigate appropriately and safely. Session cookies created by the Elettra-Sincrotrone Trieste S.C.p.A. website navigation do not affect users' privacy during their browsing experience on our website, as they do not entail processing their personal identification data. Session cookies are not permanently stored and indeed are cancelled when the connection to the Elettra-Sincrotrone Trieste S.C.p.A. website is terminated.
More info
OK

HowTo&Troubleshooting

How to switch the branchline (e.g. from A to B)

Only for beamline staff. This procedure is not inteded for users.
IMPORTANT! Before starting to move check that the value in 082 BCS: CENTRAL PEM1 HEIDENHEIM_20 ASSEZ is different from 0 and the STATUS iS 0 (AND NOT 8).
IF STATUS is 8 and encoder is 0 you need to call Michele Turcinovich and ask to reboot the ENCODER (it is not necessary to restart the bcs).
 

1. POSTHORA. Z. Read. It should be UserUnits=-23 (UPDATED GEN2011. NOTE: THE OLD VALUE WAS -26), Encoder +23 (Branchline A)

2. POSTHORB. Z. Read. It should be UserUnits=-63 (OLD VALUE WAS -66), Encoder +23 (Branchline A). If OVERWRITE or RESTORE message appears, it is likely you have to select OVERWRITE.

3. POSTHORB. Z. Move UserUnits to -15 (OLD VALUE WAS -18), Encoder XX (OLD VALUE was -22) (Branchline B).

4. POSTHORA. Z. Read. A message OVERWRITE/RESTORE will appear. Press Overwrite. Now you should read: UserUnits=+22 (THIS IS THE OLD VALUE, TO BE UPDATED), Encoder -22 (OLD VALUE, TO BE UPDATED!)(Branchline A). If OVERWRITE or RESTORE message appears, it is likely you have to select OVERWRITE.

FROM A TO C:
1. POSTHORA. Z. Read. It should be UserUnits=-23 (UPDATED GEN2011. NOTE: THE OLD VALUE WAS -26), Encoder +23 (Branchline A)
2. POSTHORA. Z. Move User Units to 0 (zero). Enc=0.

FROM C TO A
1. POSTHORA. Z. Read. It should be UserUnits=0 Encoder 0 (Branchline C)
2. POSTHORA. Z. Move User Units to -23 (zero). Enc=+23.

Before starting to move check that the value in 082 BCS: CENTRAL PEM1 HEIDENHEIM_20 ASSEZ is different from 0 and the STATUS iS 0 (AND NOT 8).
IF STATUS is 8 and encoder is 0 you need to call Michele Turcinovich and ask to reboot the ENCODER (not the bcs).
 

 