Photodiodes


The experimental chamber is equipped with 4 silicon diodes for light detection. All diodes are mounted on  a aluminum frame placed on the PhiA analyzer frame, at 160 mm from the chamber center. These detectors are all at 160 mm from the chamber center. 3 diodes are SXUV-100 model and one is AXUV-100G for absolute intensity measurements. Moreover, there is also an absolute diode AXUV-300 mounted on the monitor chamber for I0. The following table summarizes the main properties of the detectors:

Diodes properties
model distance from center (mm) mask size (mm2) V X H Acceptance (H x V) deg2 Group/connection Direct beam position
sxuv100 160 mm 8 x 8 7.6 x 7.6 A 2 ThetaA -50 PhiA 0
sxuv100 160 mm 8 x 0.5 7.6 x 0.5 B 1 ThetaA -119.6 PhiA +2
axuv100 160 mm 8 x 8 7.6 x 7.6 A 1 ThetaA -129.6 PhiA -2.5
sxuv100 160 mm 8 x 8 7.6 x 7.6 B 2 ThetaA -59.4 PhiA +2
axuv300 monitor chamber 20 x 20 ------------ A 3 ---------------------------



The diodes are connected through a DN63CF 8-ways feedthrough placed near the phiA motor.

 
Diodes mounted on PhiA frame


UHV feedthrough for photodiodes. 


Last Updated on Tuesday, 18 April 2023 11:43