ELETTRA News Number: 36 - August 31, 1999.
Phase Contrast and Diffraction Enhanced
Mammography at SYRMEP.
by F. Arfelli(1), V. Bonvicini(1), A. Bravin(2), G. Cantatore(1),
E. Castelli(1), L. Dalla Palma(3), R. Longo(1), R. H. Menk(2), A. Olivo(1),
S. Pani(1), P. Poropat(1), M. Prest(1), A. Rashevsky(1), L. Rigon(1), G. Tromba(2),
A. Vacchi(1) and E. Vallazza(1)
(1)Dipartimento di Fisica - Università di Trieste and INFN - Sez. di Trieste,
Italy
(2)Sincrotrone Trieste SCpA, Basovizza, Trieste, Italy
(3)Istituto di Radiologia dell'Università di Trieste, Ospedale di Cattinara,
Trieste, Italy
( contact: tromba@elettra.trieste.it
)
The high degree of coherence of third generation synchrotron sources such
as Elettra offers the opportunity to investigate new x-ray imaging techniques
that can be applied out also to medical radiology [1-4]. In conventional radiography
the image contrast is based on absorption effects expressed by the amplitude
properties of x-rays detected by the image receptor. Taking advantage of the
beam coherence, additional information can be obtained about the phase perturbations
of the incident wave: both spatial and contrast resolution can be significantly
enhanced. This progress becomes particularly important for the cases of weak
absorbing samples where small differences in the absorption coefficients result
in poor contrast images. In the field of medical radiology this is a characteristic
case of mammography where the recognition of low contrast nodules and/or small
calcifications inside a soft tissue matrix is needed to perform accurate breast
cancer diagnosis.
Two novel imaging techniques have been implemented at the SYRMEP beamline to
produce images of test objects and of in vitro tissue specimens enhanced
by the coherence of the beam: the PHase Contrast radiography (PHC) and the analyzer
crystal Diffraction Enhanced Imaging (DEI). For these studies images have been
taken using standard medical screen-film systems [4,5].
In the PHC radiography we are interested in the registration of the phase shifts
produced on the incident wave by the sample and the details inside it. The interference
pattern produced between the undiffracted and the diffracted (phase-shifted)
wave occurs along the border of the objects inside a very narrow angular region.
This signal can be detected if the distance between the sample and the detector
matches with the detector spatial resolution. The implementation of this technique
consists simply in selecting the sample-to-detector distance that maximize the
interference signal and scanning the sample and the detector through the beam
using two independent movement stages. The produced diffraction pattern appears
superimposed to the conventional absorption image on the film.
In the DEI set-up, an analyzer Si(1,1,1) crystal, in addition to the channel-cut
Si(1,1,1) crystal used as monochromator, is placed between the sample and the
detector to select the angular emission of x-rays outgoing the sample. This
crystal is supplied of appropriate movements to allow a fine tuning with the
beam. Two ionization chambers are used to find out the alignment of the analyzer
with respect to the monochromator, the latter is also used to evaluate the film
exposure.
The rocking curve of the two crystals is the same. Thus, when the analyzer is
aligned with the monochromator (i.e. the (1,1,1) planes of the crystals are
parallel), the former acts as a scattering rejecter: most of the photons scattered
by the sample at angles larger than the FWHM of the analyzer rocking curve are
rejected by the crystal.
If a slight misalignment between monochromator and analyzer is introduced, the
reflectivity of the analyzer is maximized for x-rays deviated of a given scattering
angle (equal to the misalignment angle). In this way it is possible to convert
the photon diffraction angles, and thus the phase
shifts, into different reflectivity coefficients.
Since the scattering angle is - to first approximation - proportional to the
gradient of the real part of the refractive index, the analyzer effectively
converts the behavior of this real part of the sample into different reflectivity
coefficients for the refracted wave, and consequently into intensity differences
on the detector. The effectiveness of these techniques can be evaluated by comparing
the images of different tissue specimens and test objects obtained at the beamline
and at a clinical mammographic unit with the same screen-film system devices
[4,5].
A comparison between a conventional radiography and a PHC image of a human breast
tissue specimen containing a calcification and some calcium deposits can be
done by observing Fig.1a and 1b.


Fig.1a - 1b: Images of a breast tissue specimen
with patological calcifications obtained with the mammographic unit (a) and
with the PHC technique (b).
The delivered mean glandular dose is the same (about 1.5 mGy) for the two images.
In the PHC image (Fig. 1b) the visibility of the details is enhanced; furthermore,
differently from the conventional image (Fig. 1a), one can realize that the
calcification is surrounded by soft tissue having a varying thickness. The density
of this tissue can be recognized from the opacity. At the right side of this
image one can also clearly see an extended vascular calcification.
The images of another human breast tissue sample, obtained at the mammographic
unit and with the DEI technique at 17 keV, are shown in Fig. 2a-2b.
Fig. 2a - 2b: Images of a breast tissue specimen
obtained with the mammographic unit (a) and with the DEI technique (b).
The conventional image (Fig. 2a) has been taken with a mean glandular dose
of 0.8 mGy. Compared to the previous one in the DEI image (Fig. 2b) the contrast
as well as the resolution are enhanced, this results in an improved definition
of the glandular component and of the visibility of calcifications. The mean
glandular dose is 0.6 mGy.
Both the imaging techniques can allow significant improvements in mammography
and do not require an increase in the delivered dose. Further investigations
have to be carried out to complete the comprehension of the DEI technique. Certainly
these novel imaging modalities will have a great impact in the near future of
medical applications when applied to in-vivo investigations.
References
[1] A.Snigirev, I.Snigireva, V.Kohn, S.Kuznetsov, I.Schelokov, Rev. Sci. Instrum.
66 (12), (1995), p. 5486-5492.
[2] D. Chapman, W.Thomlinson, R.E. Johnston, D.Washburn, E.Pisano, N.Gmur, Z.Zhong,
R.Menk, F.Arfelli, D.Sayers, Phys. Med. Biol. 42 (1997), p.2015-2025.
[3] M.Di Michiel, A.Olivo, G.Tromba, F.Arfelli, M.Assante, V.Bonvicini, A.Bravin,
G.Cantatore, E.Castelli, L.Dalla Palma, R.Longo, S.Pani, D.Pontoni, P.Poropat,
M.Prest, A.Rashevsky, A.Vacchi, E.Vallazza, Proceedings of Haga Conference 1997,
Springer-Verlag ed., p.78-82.
[4] F.Arfelli, M.Assante, V.Bonvicini, A.Bravin, G.Cantatore, E.Castelli, L.Dalla
Palma, M.Di Michiel, R.Longo, A.Olivo, S.Pani, D.Pontoni, P.Poropat, M.Prest,
A.Rashevsky, G.Tromba, A.Vacchi, E.Vallazza, F.Zanconati, Phys. Med. Biol, Vol.43,
n.10, oct. 1998, p.2845-2852.
[5] F.Arfelli, V.Bonvicini, A.Bravin, G.Cantatore, E.Castelli, L.Dalla Palma,
M.Di Michiel, M.Fabrizioli, R.Longo, R.H.Menk, A.Olivo, S.Pani, D.Pontoni, P.Poropat,
M.Prest, A.Rashevsky, M.Ratti, L.Rigon, G.Tromba, A.Vacchi, E.Vallazza, F.Zanconati,
in press on Radiology.
|