.
APE is a facility for advanced experiments on solid surfaces and nanostructured
matter. The all-UHV interconnected system includes state-of-the-art surface preparation and survey, atomic resolution scanning tunneling microscope, and two
beamlines delivering monochromatic, polarized synchrotron radiation to a high-
energy/angle resolution spectrometer and to a microscope allowing for soft-X ray
absorption and magnetic dichroism, photoemission, spin polarization mapping and
time-resolved measurements.
.
|
figure
1
|
|
|
|
|
|
layout
of the APE beamlines (up) with a blow-up of APE ‘laboratory’
(down)
click
on the figure to enlarge it
|
.
The beamlines use the synchrotron light from the Elettra storage ring.
Photons with chosen polarization are emitted simultaneously by two non-collinear
insertion devices and delivered in two distinct beamlines (Fig. 1). The line dedicated
to high-resolution photoelectron spectroscopy uses low energy photons (hn=10-100eV) (APE-LE, LE = Low photon Energy). The second line exploits photons in the
range hn=140-1500eV (APE-HE, HE = High photon Energy). The photon energy resolution after the APE-LE monochromator is better than
E/DE=15000. The first
results on APE-HE indicate E/DE=10000.
Current tests confirm negligible amount of
cross talk between the two
undulator sources. Samples can be loaded with two differentially pumped load-locks
and then transferred in UHV to any of the preparation chambers, to the STM and to
the both synchrotron radiation spectrometers (Fig. 1).
APE also allows the integration of users' specialized sample growth chambers
or modules, which will be connected to the main sample distribution chamber and
will have full access to the off-beam and on-beam facilities. Long-term users'
programs will be therefore possible, having access to beamtime based on usual assignment procedures, but having semi-permanent growth facilities and access to
the off-beam diagnostics.
.
.
beamline
The variable-polarization light is delivered from the Apple II quasi-periodic undulator.
It is then deviated onto the plane grating monochromator by the spherical mirror in
the first mirror chamber (Fig. 1). Three variable-spacing gratings, 700 l/mm, 1200
l/mm and 1600 l/mm, cover the energy range 10-25 eV, 25-40 eV and 40-100 eV, respectively. Currently 20-100 eV range is available since the delivery of the 700
l/mm grating is still pending. The monochromatic light is focused by a spherical
mirror on the exit slit.
The photon resolution after the exit slit is determined by measuring absorption
spectra of rare gases (Fig. 2). We find E/DE = 16000 at
hn~47 eV (Fig. 2a) and E/DE
= 13000 at hn~63 eV photons for all polarizations (Fig. 2b and 2c). The photon flux at
best resolution is 2·1010 photons/s. The stability of the photon energy is such that
we find <0.5 meV shift during 7 hours of measurements. The toroidal mirror refocuses the light further into the LE end station on 50x100
mm spot.
.
|
figure
2
|
|
 |
|
Ne (a) and He (b and c) absorption spectra used to determine the photon
energy resolution after the APE-LE monochromator
|
.
end
station
The APE-LE end station is equipped with Scienta SES-2002 electron energy
analyzer. The experimental resolution (i.e. combined light + analyzer + sample
temperature) is determined from the width of the measured Fermi edge of a polycrystalline Ag sample. We consistently measure 6 meV (Fig. 3) for 25-47 eV
photons.
The angular acceptance of the analyzer is ~14°. The angular resolution of the
analyzer is achieved electronically, with a special set of lens parameters, which
permit that all electrons leaving the sample at a given angle are imaged to the same
position at the exit slit of the analyzer. It is estimated to be <0.4°.
.
|
figure
3 |
|
 |
|
fermi edge measured on a polycrystalline
Ag sample with hn =
47 eV |
.
Depending on the type of the experiment, two alternative manipulators are
available at APE-LE. A Rial translator, with x, y, z, Q
degrees of movement, hosts a homemade cryostat (University of Zürich, Osterwalder group) that reaches ~15 K on
the sample surface. A VG translator contains the automated manipulator (built at the University of Zürich, Osterwalder group) with the x, y, z,
Q, f degrees of movement. It
is optimized for the automatic Fermi surface mapping (Figure 4). The lowest temperature for the Fermi surface mapping is ~100 K.
.
|
figure
4 |
|

|
|
measured
fermi surface of Ag(100);
hn
= 25 eV, measuring time ~7 hours |
.
This end station is also equipped with a He discharge lamp, reverse-LEED
and Ar sputter gun. There is a possibility for introducing compact evaporators. More
sample treatments can be done in the annex preparation chamber(s).
.
.
beamline
Similar to LE, the light from a periodic Apple II undulator is deviated by a spherical
mirror on the plane grating monochromator which covers 140-1500 eV energy range
(900, 1400, 1800 l/mm variable spacing gratings). The spherical mirror focuses the
monochromatic light on the exit slit. Only a first test grating (900 l/mm) has been delivered and installed insofar. The photon resolution after the exit slit is determined
by measuring the N2 and Ne absorption spectra. We find E/DE = 10000 at
hn~400 eV (Fig. 5). However, E/DE = 3000 at
hn~900 eV photons. This reduced resolution at higher photon energies is the consequence of the grating fabrication problems and is
going to be improved with the new set of gratings whose delivery is pending. The
photon flux at best resolution is 2·1010 photons/s.
.
|
figure
5 |
|
 |
|
N2 absorption spectra used to determine the photon
energy resolution after the APE-HE monochromator |
.
Two focusing schemes can be alternatively chosen when operating APE-HE:
single reflection focussing by a toroidal mirror that produces a spot size of 25·100
mm, or deviation by a plane mirror onto a Fresnel zone plate diffractive optics. The
spot size in this case is in the range 50-200 nm. The sample is moved in the wanted
position by stepper motors with XY scanning capability (0.5 mm resolution). The
microscopy is performed by scanning the Fresnel zone plate by piezoelectric drives.
.
end
station
The APE-HE end station will work as a highly focused soft-X ray spectrometer
for core level photoemission, time resolved surface magnetometry and magnetic microscopy. The chamber hosts an Omicron 125 mm analyzer for photoelectron
spectro(micro)scopy. A total yield Mott detector will be used for the surface magnetometry. The Mott detector, built at the Technical University of St. Petersburg,
is under test with the counting electronics. Special electronics for time resolution
mode (<100ps) is under development. APE HE is exploiting X-rays with variable polarization to study the magnetic
properties of materials. A test experiment was lately performed on an iron thin film
epitaxially grown on GaAs(001)-(6x4). A Fe0.5Co0.5 probing layer was deposited
between two 6 ML's-thick Fe wedges to test the magnetic properties of the film as a
function of distance from the interface with the semiconductor (see Fig. 6,
upper part). This
was done using X-Ray magnetic circular dichroism (XMCD) on the L2,3 threshold of
Co. In Fig. 6, lower part we show two spectra taken with the X-ray beam shining on the probing layer at different distances from the interface. The reducing of the dichroic
intensity is a sign that the magnetization film is reduced when approaching the
interface.
.
|
figure
6 |
|

|
 |
|
upper part: a
Fe0.5Co0.5 probing layer is sandwiched between two wedges of bcc Fe
deposited on GaAs(001)-6x4;
lower part: XMCD spectra taken at different probing distances from the
metal/semi-
conductor interface |
.
|
ape
preparation and stm chambers |
|
|
|
|
.
APE has two sample preparation stages. One of them was developed at the
ETH Zürich (Pescia group) and transferred to Trieste in February 2000. It is equipped with LEED/Auger system, for both the chemical and structural
characterization of the sample surfaces; Ar+ sputter gun; high temperature heating
stage (2500°C); evaporators for epitaxial growth of metal thin films; room temperature scanning tunneling microscope (see Fig. 7) with in-situ tip and sample
exchange. In addition, magnetic monolayers and multilayer systems can be investigated with a vectorial magneto-optical Kerr effect. The second preparation
stage is located in the linear transfer chamber attached to the APE-LE end station. It
is designed to host an Ar+ sputter gun, three evaporators and a sample heating stage.
.
|
figure
7 |
|
 |
|
atomically resolved
stm image for sulfur
segregated Fe(3%Si) (100); the zigzag
lines along <110> appear on c(2x2)-S;
82x82Å (+400mV, 0.1nA) |
.
Samples and substrates can be loaded in the APE system via either one of
the two identical load-locks connected to the STM chamber and to the switching chamber between the UFO and the linear travel chamber. The sample preparation
can be done in either of the two preparation stages. Substrates or samples can be
introduced also from users' sample chambers docked at one of the two dedicated flanges in the UFO chamber (100 mm CF flanges indicated in Fig. 1). Through those
ports the arm of the UFO can enter and collect the sample that can be measured by
any of the characterization tools and finally exposed to synchrotron radiation for
experiments. All interested users can ask for a design of the UFO chamber and the
environment for evaluating the possibility of using their system to APE.
.
.
APE beamline received first official users' proposals in August 2002. In the period
January-June 2003 the full commissioning continues in parallel with the experiments
on the basis of 20% of the available beamtime. From July 2003 APE is fully
operational for users.
As the delivery of the pending elements (diffraction gratings, electron transport lenses for spin polarized measurements, electronics for time resolution
pump-probe experiments, cryostats) will take place, the full deployment of the APE
project will be pursued.
.
Prof. Giorgio Rossi, head (giorgio.rossi@tasc.infm.it)
Dr. Giancarlo Panaccione, senior beamline scientist (panaccioneg@elettra.trieste.it)
Dr. Ivana Vobornik, beamline scientist (ivana.vobornik@elettra.trieste.it)
Dr. Jun Fujii, beamline scientist (jun.fujii@elettra.trieste.it)
Dr. Michael Hochstrasser, beamline scientist, ETH Zürich (CH) (michael.hochstrasser@elettra.trieste.it)
Dr. Luca Giovanelli, post-doc fellow (luca.giovanelli@elettra.trieste.it)
Damjan Krizmancic, software engineer (damjan.krizmancic@elettra.trieste.it)
Francesco Maccherozzi, PhD student
Mauro Fabrizioli, PhD student
Matia Mulazzi, PhD student
.
Dr. Jörg Kröger, now at University of Kiel (D)
Dr. Stefan Egger, now at University of Cambridge (UK)
Dr. Cinzia Cepek, now at TASC Laboratory, Trieste (I)
.
Prof. Christian Back, University of Regensburg (D)
Urs Maier, ETH Zürich (CH)
Prof. Silvio Modesti, University of Trieste (I)
Prof. Jürg Osterwalder, University of Zürich (CH)
Prof. Danilo Pescia, ETH Zürich (CH)
Dr. Francois Polack, LURE (F)
Prof. Massimo Sancrotti, University of Brescia (I)
Dr. Andreas Vaterlaus, ETH Zürich (CH)