Elettra-Sincrotrone Trieste S.C.p.A. website uses session cookies which are required for users to navigate appropriately and safely. Session cookies created by the Elettra-Sincrotrone Trieste S.C.p.A. website navigation do not affect users' privacy during their browsing experience on our website, as they do not entail processing their personal identification data. Session cookies are not permanently stored and indeed are cancelled when the connection to the Elettra-Sincrotrone Trieste S.C.p.A. website is terminated.
More info
OK

Specifications


Spatial resolution
The microprobe is formed using Schwartzschild objectives with multilayers. Two can be installed in UHV at the same time. We have available 27 eV, 74 eV and 95 eV objectives.

   Flux and line width on the sample  (monochromator slit 100 µm)
 
27 eV 2x1010 photons sec -1 5 meV
74 eV 1.5x1011 photons sec -1 70 meV
95 eV 2x1011 photons sec -1 100 meV

    Long range background

We use mostly 27 eV with less than 1 % Background and 74 eV with less than 25 % background.

   Microprobe size

 

    Focal depth

Because of large (0.2) numeric apertire the focal depth is small. We need to focus with precision of approximately 2 µm.