Elettra-Sincrotrone Trieste S.C.p.A. website uses session cookies which are required for users to navigate appropriately and safely. Session cookies created by the Elettra-Sincrotrone Trieste S.C.p.A. website navigation do not affect users' privacy during their browsing experience on our website, as they do not entail processing their personal identification data. Session cookies are not permanently stored and indeed are cancelled when the connection to the Elettra-Sincrotrone Trieste S.C.p.A. website is terminated.
More info
OK

Specifications


Temporal resolution
This feature is currently under development. First tests showed that if the pass energy of the analyzer is above 10 V we can attribute from which bunch the photoelectrons originate thereby allowing single bunch (tens of psec) resolution in time resolved exeperiments even in multibunch Elettra mode.

     Bunch structure

The bunch width seen by the detector is approximately 1 nsec depending on analyzer pass energy and lens mode and is due to photoelectron trajectories spread caused by finite slit width of the analyzer, however photoelectrons in the same bunch originate from the sampe in the same time moment within bunch lens resolution (tens of psec).

    Time structure on the detector

On the image detector structure for the electrons filtered in a window 100 ps is shown. X axis is energy, Y axis is angle.
Six bunches arrive contemporarily and the zig-zag structure is caused by  imperfections of the entrance slit of the analyzer (to be impropved).