Elettra-Sincrotrone Trieste S.C.p.A. website uses session cookies which are required for users to navigate appropriately and safely. Session cookies created by the Elettra-Sincrotrone Trieste S.C.p.A. website navigation do not affect users' privacy during their browsing experience on our website, as they do not entail processing their personal identification data. Session cookies are not permanently stored and indeed are cancelled when the connection to the Elettra-Sincrotrone Trieste S.C.p.A. website is terminated.
More info
OK

Specifications


Angular and Energy resolution

   Energy resolution (currently the analyzer entrance slit size is 500 µm along energy dispersion and 12 mm along angle dispersion)

27 eV photons (5 o 10 meV line width)
Pass Energy (V) Resolution (meV)
5 24
10 47
20 93

74 eV photons
Pass Energy (V) Resolution (meV)
monochromator slit 50 µm
Resolution (meV)
monochromator slit 100 µm
5 50 90
10 65 100
20 102 130

   Angular resolution


Analyzer mode Resolution along angular dispersion Resolution along energy dispersion Angle accepted along angular dispersion
WA ±0.27o ±0.67o ±8o
AP0.75 ±0.13o ±0.33o ±8o
AP1 ±0.1o ±0.25o ±6o
AP1.5 ±0.07o ±0.17o ±4o
AP2 ±0.05o ±0.13o ±3o

Last Updated on Thursday, 08 March 2012 14:49